User Manual
COMMAND OVERVIEW
183
RO (output impedance) is setting 25Ω, the setup value of AC
voltage is {0.01~1|MAXimum|MINimum}.
Set syntax
:MEASure:VOLTage:AC 1
:MEASure:VOLTage:AC 1M
:MEASure:VOLTage:AC 1MV
:MEASure:VOLTage:AC 1E-3
:MEASure:VOLTage:AC MAXimum
:MEASure:VOLTage:AC MINimum
Query syntax
:MEASure:VOLTage:AC?
Return data 1.000000E+00 (Format is in <NR3>)
When the Level setup is current mode, the data returns 9.9E37.
:MEASure:CURRent:AC <current NR3/disc>
:MEASure:CURRent:AC?
Function: It sets or queries the AC measurement current (LEVEL).
Description:
Set parameter When RO(output impedance) is setting 100Ω, the
setup value of AC current is {0.0002~0.02|MAXimum|MINimum}.
When RO (output impedance) is setting 25Ω, the setup value of AC
current is {0.0002~0.04|MAXimum|MINimum}.
Set syntax
:MEASure:CURRent:AC 0.01
:MEASure:CURRent:AC 10M
:MEASure:CURRent:AC 10MA
:MEASure:CURRent:AC 1E-2
:MEASure:CURRent:AC MAXimum
:MEASure:CURRent:ACMINimum
Query syntax
:MEASure:CURRent:AC?
Return data 1.000000E-02 (Format is in <NR3>)
When the Level setup is voltage mode, the data returns 9.9E37.