User Manual
CORRECTION (OPEN/SHORT)
137
Setting reference
Input the real part and imaginary part value with the part .
Steps
1. Press the Measure button to enter
[MEAS DISPLAY] page, and then
press the CORR. function button on
the right side of the LCD to enter the
[CORRECTION] page.
2. Use arrow keys to move cursor and
select REFERENCE item on the
[CORRECTION] page.
3. Use the numeric keypad to enter the real part
(Ls) and imaginary part (Q) value.