User Manual

LCR-8000 Series User Manual
128
Note
If HF load correction fails, you should check the
following points before starting the correction
process again:
Check that the test fixture or the test cables are
properly connected.
Check that the test fixture or the test cables are
properly shorted together with the STD-
LOAD.
During the correction process, be sure not to
disturb the fixture and the test cables or to
move your hand near them.
STD-LOAD
(LCR-05A,100Ω)