CORRECTION (OPEN/SHORT)
127
Setting HF load
This item does not necessarily need to be implemented. Some test
fixtures have a standard LOAD, which can be performed with
reference to the fixture instructions and requirements to perform
LOAD (PHASE) compensation.
Steps
1. Press the Measure button to enter
[MEAS DISPLAY] page, and then
press the CORR. function button on
the right side of the LCD to enter the
[CORRECTION] page.
2. Use arrow keys to move cursor and
select HF LOAD item on the
[CORRECTION] page.
3. Use option key on the right of the LCD screen
to select HF LOAD item to start execution.
4. When the LOAD compensation is completed,
the HF LOAD status will automatically switch
to ON.