User Manual

LCR-8000 Series User Manual
104
Setting OUTPUT IMPEDANCE
The output impedance can be set to 2 or 100Ω.
The varied signal source output impedance will lead to the varied
current or the difference of measuring value. If selecting <25Ω>,
then the voltage range is 10mV~1Vrms and the current range is
400µA~40mArms.
If need to compare test results with Keysight select 100Ω.
Steps
1. Press the Measure button to enter
[MEAS DISPLAY] page, and press the
LIST SET function button on the right
side of the LCD to enter the [LIST SET]
page, and then press the Setup button
to enter [LIST MODE SETUP] page.
2. Use arrow keys to move cursor and
select OUTPUT IMPEDANCE item on
the [LIST MODE SETUP] page.
3. Use option key on the right of the LCD screen to
select an impedance item.
Available
parameter
100Ω
Set output impedance 100Ω
25Ω
Set output impedance 25Ω