User Manual
LIST (MULTI STEP MODE)
101
Available
parameter
REPEAT
The repeat trigger mode is an internal
continuous test. Press the Trigger
button to start all the steps of the LIST
test, and repeat all the steps until the
STOP function button on the right side
of the LCD is pressed.
SINGLE
External trigger mode, including
Manual/Handler/TRIGGER
Input/Remote control mode.
Manual mode: The device performs
a measurement once the Trigger
button is pressed.
Handler mode: When a negative
edge pulse is received from the
handler interface on the rear panel,
the device performs a measurement
cycle.
TRIGGER Input mode: When a
negative edge pulse is received
from the TRIGGER Input on the
rear panel, the device performs a
measurement cycle.
Remote control mode: When a
measurement command is sent
from the RS-232 or USB or GPIB
interface, the device performs a
measurement cycle.
AUTO
Automatic trigger mode.
After the measurement terminal is
open, it will enter the standby mode
"Wait On" , the instrument
will always measure, after measuring
the value, it will start the trigger
condition for measurement, and after
the LIST step test, it will enter the
waiting for end mode "Wait