User manual
19
z Off Line
Function: PC disconnect with LCR Meter. PC have to send a command COMU:OFF.^END^M( or ^J^M )to LCR Meter. LCR Meter will
respond a command COMU:OFF.^END to PC and recover display after receive PC command.
Command String:
PC sends command COMU:OFF.< ^END^M> or <^J^M>
ASCII CODE = 43 4F 4D 55 3A 4F 46 46 2E 0A 0D (Hex format)
LCR Meter response: COMU:OFF.<^END>