User manual

8
Set Nomval
PC send command SORT:NOMV +32.0000< ^END^M> or <^J^M>
ASCII CODE = 53 4F 52 54 3A 4E 4F 4D 56 20 2B 33 32 2E 30 30 30 30 0A 0D (Hex format)
LCR Meter Response SORT:NOMV 32.0000<^END>
ASCII CODE = 53 4F 52 54 3A 4E 4F 4D 56 20 20 33 32 2E 30 30 30 30 0A (Hex format)
PC send command SORT:NOMV -32.0000< ^END^M> or <^J^M>
ASCII CODE = 53 4F 52 54 3A 4E 4F 4D 56 20 2D 33 32 2E 30 30 30 30 0A 0D (Hex format)
LCR Meter Response SORT:NOMV -32.0000<^END>
ASCII CODE = 53 4F 52 54 3A 4E 4F 4D 56 20 2D 33 32 2E 30 30 30 30 0A (Hex format)
Set Voltage
PC send command MAIN:VOLT 1.000< ^END^M> or <^J^M>
ASCII CODE = 4D 41 49 4E 3A 56 4F 4C 54 20 31 2E 30 30 30 0A 0D (Hex format)
LCR Meter Response MAIN:VOLT 1.000<^END>
ASCII CODE = 4D 41 49 4E 3A 56 4F 4C 54 20 31 2E 30 30 30 0A (Hex format)
Set Average
PC send command STEP:AVER 1.00< ^END^M> or <^J^M>
ASCII CODE = 53 54 45 50 3A 41 56 45 52 20 31 2E 30 30 0A 0D (Hex format)
LCR Meter Response STEP:AVER 1.00<^END>
ASCII CODE = 53 54 45 50 3A 41 56 45 52 20 31 2E 30 30 0A (Hex format)
Store Memory
PC send command MEMO:STOR 1.00< ^END^M> or <^J^M>