User manual
7
Example:
ONLINE step 1(power on only)
PC send command COMU?< ^END^M> or <^J^M>
ASCII CODE = 43 4F 4D 55 3F 0A 0D (Hex format)
LCR Meter Response : COMU:ON..<^END>
ASCII CODE = 43 4F 4D 55 3A 4F 4E 2E 2E 0A (Hex format)
ONLINE step 2(power on only, LCR METER WILL SHOW RS232 ONLINE)
PC send command COMU:OVER< ^END^M> or <^J^M>
ASCII CODE = 43 4F 4D 55 3A 4F 56 45 52 0A 0D (Hex format)
LCR Meter Response : COMU:OVER<^END
ASCII CODE = 43 4F 4D 55 3A 4F 56 45 52 0A (Hex format)
Set Speed
PC send command MAIN:SPEE:FAST< ^END^M> or <^J^M>
ASCII CODE = 4D 41 49 4E 3A 55 50 45 45 3A 46 41 53 54 0A 0D (Hex format)
LCR Meter Response : MAIN:SPEE:FAST<^END>
ASCII CODE = 4D 41 49 4E 3A 53 50 45 45 3A 46 41 53 54 0A (Hex format)
Set Frequency
PC send command MAIN:FREQ 1.00000< ^END^M> or <^J^M>
ASCII CODE = 4D 41 49 4E 3A 46 52 45 51 20 31 2E 30 30 30 30 30 0A 0D (Hex format)
LCR Meter Response: MAIN:FREQ 1.00000<^END>
ASCII CODE = 4D 41 49 4E 3A 46 52 45 51 20 31 2E 30 30 30 30 30 0A (Hex format)