User manual
PROGRAMMING
103
AVERAGE
Command/Query
Sets the average number from 1~255. The average number indicates
how many test samples are used to create an averaged test result.
Syntax S
TEP:AVER <variable><^END^M>or<^J^M>
Parameter
<variable> Average number
1.00~255. (integer values) 1~255
Note
Ensure the number has a total of 4 characters. If a
number does not use 4 characters, use a “.” and
additional zero’s (0) to “pad out” the number.
Example 10 = 10.0
Example STEP:AVER 255.<^END^M>
Average is set to 255 samples.
Query Syntax S
TEP:AVER?<^END^M>or<^J^M>
Return String
<string> Current average setting
STEP:AVER <variable><^END>
(<variable>= 1.00 ~255.)
Returns the average
number.
Query Example
STEP:AVER?<^END^M>
STEP:AVER 255.<^END>
The average number is currently 255.
RECALL CALIBRATION
Command
Recalls the calibration settings from memory. A return string
indicates if the command was successful.
Syntax STEP:RECA<^END^M>or<^J^M>
Return String
<string> Recall calibration
RECA:OK<^END>
Successful
LCR-800 User Manual
104
Example
STEP:RECA<^END^M>or<^J^M>
RECA:OK<^END>
Calibration was recalled successfully
BAUD RATE
Command
Sets the baud rate of the RS232 connection.
Syntax COMU:<value><^END^M>or<^J^M>
Parameter
<value> Baud rate
9600 9600
19.2 19200
38.4 38400
57.6 57600
1152 115200
Return String
<string> Baud rate
COMU:<value><^END>
<value>= baud rate
Returns the baud rate
setting.
Query Example
COMU:1152<^END^M>
COMU:1152<^END>
The baud rate is set to 115200.
MODEL NUMBER
Query
This query returns the model number of the LCR-800.
Query Syntax COMU:MON
O<^END^M>or<^J^M>
Return String
<string> Model number
COMU:MONO:816.<^END>
LCR-816
COMU:MONO:817.<^END>
LCR-817