User manual
GSP-9300 Programming Manual
88
:CALCulate:DEMod:AM:RESult:MINimum?
Query
Description
Returns the minimum recorded measurement
results for AM demodulation as a comma
separated string.
Query Syntax
:CALCulate:DEMod:AM:RESult:MINimum?
Return parameter
<depth,rate,power,offset,sinad>
depth
Modulation depth in %.
rate
Modulation rate in Hz.
power
Carrier power in the current Y-axis units.
offset
Carrier frequency offset in Hz.
sinad
Signal to noise and distortion ratio in dB
Query Example
:CALC:DEM:AM:RES:MIN?
>9.840e+1,1.02e+2,-1.12e+1,3.21e+1,1.61e+1
:CALCulate:DEMod:AM:RESult:MAXimum?
Query
Description
Returns the maximum recorded measurement
results for AM demodulation as a comma
separated string.
Query Syntax
:CALCulate:DEMod:AM:RESult:MAXimum?
Return parameter
<depth,rate,power,offset,sinad>
depth
Modulation depth in %.
rate
Modulation rate in Hz.
power
Carrier power in the current Y-axis units.
offset
Carrier frequency offset in Hz.
sinad
Signal to noise and distortion ratio in dB.
Query Example
:CALC:DEM:AM:RES:MAX?
>9.840e+1,1.02e+2,-1.12e+1,3.21e+1,1.61e+1
:CALCulate:DEMod:AM:STATe
Set
Query
Description
Sets or queries the state of the AM Analysis
function.