User manual
GSP-9300 Programming Manual
182
[:SENSe]:SEMask:OFFSet<n>:STOP:
RELative:COUPle
Set
Query
Description
Couples the Relative Stop amplitude to the
Relative Start amplitude for the selected offset.
Syntax
[:SENSe]:SEMask:OFFSet<n>:STOP:RELative:COUPle
{OFF|ON|0|1}
Query Syntax
[:SENSe]:SEMask:OFFSet<n>:STOP:RELative:
COUPle?
Parameter
0
Turns coupling off.
1
Turns coupling on.
OFF
Turns coupling off.
ON
Turns coupling on.
Return parameter
0
Coupling is off.
1
Coupling is on.
[:SENSe]:SEMask:OFFSet<n>:TEST
Set
Query
Description
Sets or queries the masks to use for the Fail
Mask(s).
Syntax
[:SENSe]:SEMask:OFFSet<n>:TEST
{ABS|REL|AND|OR}
Query Syntax
[:SENSe]:SEMask:OFFSet<n>:TEST?
Parameter/
Return parameter
ABS
Absolute mask
REL
Relative mask
AND
Absolute and Relative mask
OR
Absolute or Relative mask
Example
:SEM:OFFS:1:TEST ABS
[:SENSe]:SEMask:SELect
Set
Query
Description
Selects or queries the type of spectrum emission
mask.