User manual
GSP-9300 Programming Manual
150
Return parameter
0
Autoset amplitude floor is in manual.
1
Autoset amplitude floor is in auto.
Example
:ASET:AMPL:AUTO 1
[:SENSe]:ASET:RUN
Set
Description
Activates the Autoset function.
Syntax
[:SENSe]:ASET:RUN
Example
:ASET:RUN
[:SENSe]:ASET:SPAN
Set
Query
Description
Sets or queries the Autoset span.
Syntax
[:SENSe]:ASET:SPAN <freq>
Query Syntax
[:SENSe]:ASET:SPAN?
Parameter
<freq>
<NRf>
Return parameter
<NR3>
Hz
Example
ASET:SPAN 2.0e+6
[:SENSe]:ASET:SPAN:AUTO
Set
Query
Description
Turns the Autoset span to auto or manual or
queries its state.
Syntax
[:SENSe]:ASET:SPAN:AUTO {OFF|ON|0|1}
Query Syntax
[:SENSe]:ASET:SPAN:AUTO?
Parameter
0
Turn Autoset span to manual (off).
1
Turn Autoset span to automatic (on).
OFF
Turn Autoset span to manual (off).
ON
Turn Autoset span to automatic (on).
Return parameter
0
Autoset span is set to manual (off).
1
Autoset span is set to automatic (on).