Return parameter
<freq>
Frequency resolution in Hz.
Note
* Only one marker can be selected at a time to use the
marker counter function.
The selected marker counter will disable the
previously selected marker counter.
** Only 1000, 100, 10, 1 Hz are meaningful.
Example
:CALC:MARK1:FCO:RES?
>1.0e+3
:CALCulate:MARKer<n>:FCOunt:
RESolution:AUTO
Set
Query
Description
Sets the frequency counter resolution Auto setting
on/off.
Syntax
:CALCulate:MARKer<n>:FCOunt:RESolution:AUTO
{ON|OFF|1|0}
Query Syntax
:CALCulate:MARKer<n>:FCOunt:RESolution:AUTO?
Parameter
<n>
<NR1>Marker number 1~6
0
Auto is off.
1
Auto is on.
OFF
Auto is off.
ON
Auto is on.
Return parameter
0
Auto is off.
1
Auto is on.
Example
:CALC:MARK1:FCO:RES:AUTO?
:CALCulate:MARKer<n>:FCOunt:STATe
Set
Query
Description
Sets or queries the state of the frequency counter
function.
Syntax
:CALCulate:MARKer<n>:FCOunt:STATe {ON|OFF|1|0}
Query Syntax
:CALCulate:MARKer<n>:FCOunt:STATe?