User manual

GSP-9300 Quick Start Guide
52
Spurious Response
Second Harmonic
Intercept
Preamp off; signal input -30dBm; 0
dB attenuation
+35 dBm
Typical
; 10 MHz < fc < 775 MHz
+60 dBm
Typical
; 775 MHz fc < 1.5 GHz
Third-order
Intercept
Preamp off; signal input -30dBm; 0
dB attenuation
> 1dBm
300 MHz to 3 GHz
Input Related
Spurious
< -60 dBc
Input signal level -30 dBm, Att.
Mode, Att=0dB; 20-30ºC
Residual
Response
(inherent)
<-90 dBm
Input terminated; 0 dB attenuation;
Preamp off
Sweep
Sweep Time
Range
310 us to 1000 s
Span > 0 Hz
50 us to 1000 s
Span = 0 Hz; Min
Resolution = 10 us
Sweep Mode
Continuous; Single
Trigger Source
Free run; Video; External
Trigger Slope
Positive or negative edge
RF Preamplifier
Frequency Range
1 MHz to 3 GHz
Gain
18 dB
Nominal
(installed as
standard)
Front Panel Input/Output
RF Input
Connector Type
N-type female
Impedance
50 ohm
Nominal
VSWR
<1.6 :1
300 kHz to 3 GHz; Input
attenuator 10 dB