User manual
REMOTE CONTROL
91
:CALCulate:DEMod:FM:STATe
Set
Query
Description
Sets or queries the state of the FM Analysis
function.
Syntax
:CALCulate:DEMod:FM:STATe {ON|OFF|1|0}
Query Syntax
:CALCulate:DEMod:FM:STATe?
Parameter
0
Turn FM Analysis off.
1
Turn FM Analysis on.
OFF
Turn FM Analysis off.
ON
Turn FM Analysis on.
Return parameter
0
FM Analysis is off.
1
FM Analysis is on.
Example
:CALC:DEM:FM:STAT 1
:CALCulate:DEMod:RESet
Set
Description
This command will reset the max and min records
for the current demodulation analysis.
Syntax
:CALCulate:DEMod:RESet
:CALCulate:HARMonic:DISTortion?
Query
Description
Returns the harmonic distortion as a percentage of
the fundamental and as dBc.
Query Syntax
CALCulate:HARMonic:DISTortion?
Return parameter
<%>,<dBc>
<%>
THD as %. <NR1> format
<dBc>
THD as dBc. <NRf> format
Query Example
CALC:HARM:DIST?
>32.34,-9.81e+00