User manual
REMOTE CONTROL
81
Parameter
<n>
<NR1>adjacent channel 1~3
Return parameter
<NR3>
Power ratio in dB
Query Example
:CALC:ACPR:ACH1:UPP?
>1.921e+01
:CALCulate:ACPR:ACHannel<n>:STATe
Set
Query
Description
Sets or queries the state of the selected adjacent
channel.
Syntax
:CALCulate:ACPR:ACHannel<n>:STATe {OFF|ON|0|1}
Query Syntax
:CALCulate:ACPR:ACHannel<n>:STATe?
Parameter
<n>
<NR1>adjacent channel 1~3
0
Disable the selected channel.
1
Enable the selected channel.
OFF
Disable the selected channel.
ON
Enable the selected channel.
Return parameter
0
The selected channel is disabled.
1
The selected channel is enabled.
Query Example
:CALC:ACPR:ACH1:STAT?
>1
:CALCulate:ACPR:CHANnel:HLIMit:FAIL?
Query
Description
Returns the ACPR upper limit pass/fail judgment
for the main channel. A pass indicates that every
trace point in the main channel is lower than or
equal to the uppder limit.
Query Syntax
:CALCulate:ACPR:CHANnel:HLIMit:FAIL?
Return parameter
0
<boolean>Pass
1
<boolean>Fail
Query Example
:CALC:ACPR:CHAN:HLIM:FAIL?
>0