REMOTE CONTROL
171
Query Syntax
[:SENSe]:SEMask:BANDwidth|BWIDth[:RESolution]:
AUTO?
Parameter
0
Turn RBW to manual (off).
1
Turn RBW to auto (on).
OFF
Turn RBW to manual (off).
ON
Turn RBW to auto (on).
Return parameter
0
RBW is set to manual.
1
RBW is set to automatic.
Example
:SEM:BAND: AUTO OFF
[:SENSe]:SEMask:CARRier:AUTO
Set
Query
Description
Turns the PSDRef or TotalPwrRef modes to auto
(on) or manual (off) for SEM measurements or
queries their state.
Syntax
[:SENSe]:SEMask:CARRier:AUTO {OFF|ON|0|1}
Query Syntax
[:SENSe]:SEMask:CARRier:AUTO?
Parameter
0
Turn PSDRef/TotalPwrRef to manual
(off).
1
Turn PSDRef/TotalPwrRef to auto (on).
OFF
Turn PSDRef/TotalPwrRef to manual
(off).
ON
Turn PSDRef/TotalPwrRef to auto (on).
Return parameter
0
PSDRef/TotalPwrRef is set to manual.
1
PSDRef/TotalPwrRef is set to automatic.
Example
:SEM:CARR:AUTO OFF
[:SENSe]:SEMask:CARRier:CPSD
Set
Query
Description
Sets or queries the value of the PSDRef for SEM
measurements.
Syntax
[:SENSe]:SEMask:CARRier:CPSD <NR3>
Query Syntax
[:SENSe]:SEMask:CARRier:CPSD?