User manual
REMOTE CONTROL
167
Parameter
<rel_ampl>
<NRf>
Return parameter
<NR3>
Returns the gain offset value in dB.
Example
:P1DB:AVER:COUN 10.00e+00
[:SENSe]:PMETer:FREQuency
Set
Query
Description
Sets or queries the power meter measurement
frequency.
Syntax
[:SENSe]:PMETer:FREQuency <freq>
Query Syntax
[:SENSe]:PMETer:FREQuency?
Parameter
<freq>
<NRf>
Return parameter
<NR3>
Hz
Example
:PMET:FREQ 2e+7
[:SENSe]:PMETer:HLIMit
Set
Query
Description
Sets or queries the power meter high limit for
pass/fail tests.
Syntax
[:SENSe]:PMETer:HLIMit <ampl>
Query Syntax
[:SENSe]:PMETer:HLIMit?
Parameter
<ampl>
<NRf> power unit, default = dBm
Return parameter
<NR3>
Unit = current unit.
Example
:PMET:HLIM 10
[:SENSe]:PMETer:HOLD:STATe
Set
Query
Description
Turns the power meter Max/Min Hold function
on/off or queries its state.
Syntax
[:SENSe]:PMETer:HOLD:STATe {OFF|ON|0|1}
Query Syntax
[:SENSe]:PMETer:HOLD:STATe?