User manual
REMOTE CONTROL
159
Parameter
<freq>
<NRf>
Return parameter
<NR3>
Hz
Example
DEM:IFBW 3.0e+5
[:SENSe]:DEMod:SQUelch:LEVel
Set
Query
Description
Sets or queries the carrier squelch level.
Syntax
[:SENSe]:DEMod:SQUelch:LEVel <dBm level>
Query Syntax
[:SENSe]:DEMod:SQUelch:LEVel?
Parameter
<dBm level>
<NRf>
Return parameter
<NR3>
Squelch level in dBm
Example
DEM:SQU:LEV 1.30e+2
[:SENSe]:DETector[:FUNCtion]
Set
Query
Description
Sets or queries the trace detection mode when in
manual mode.
Syntax
[:SENSe]:DETector[:FUNCtion]
{AVERage|SAMPle|POSitive|NEGative|NORMal}
Query Syntax
[:SENSe]:DETector[:FUNCtion]?
Parameter/
Return parameter
AVERage
Sets the detector mode to Average.
SAMPle
Sets the detector mode to Sample.
POSitive
Sets the detector mode to Peak+.
NEGative
Sets the detector mode to Peak-.
NORMal
Sets the detector mode to Normal.
Example
:DET NORM
[:SENSe]:DETector[:FUNCtion]:AUTO
Set
Query
Description
Turns the trace detection mode to auto (on) or
manual (off) or queries its state.
Syntax
[:SENSe]:DETector[:FUNCtion]:AUTO {OFF|ON|0|1}