User manual
REMOTE CONTROL
125
:DISPlay[:WINDow]:TRACe<n>:MODE
Set
Description
Sets the operation mode of the selected trace.
Syntax
:DISPlay[:WINDow]:TRACe<n>:MODE
{WRITe|VIEW|BLANk|MAXHold|MINHold}
Parameter
<n>
<NR1> Trace number 1~4
WRITe
Clear and Write
VIEW
Hold the last trace
BLANK
Clears the trace
MAXHold
Hold the maximum/minimum points
from each sweep
MINHold
Example
:DISP:TRAC4:MODE VIEW
:DISPlay[:WINDow]:TRACE<n>:MODE:MAX
HOLD?
Query
Description
Returns the maxhold threshold level. Any part of
the trace below this threshold won’t be held when
the detector is set to Maxhold.
Query syntax
:DISPlay[:WINDow]:TRACE<n>:MODE:MAXHOLD?
Return parameter
<n>
<NR1> Trace number.
<NR3>
Example
:DISP:TRACE2:MODE:MAXHOLD?
>-2.000e+01
:DISPlay[:WINDow]:TRACe:Y:DLINe
Set
Query
Description
Sets or queries the display line amplitude level.
Syntax
:DISPlay[:WINDow]:TRACe:Y:DLINe <ampl>
Query Syntax
:DISPlay[:WINDow]:TRACe:Y:DLINe?