User manual
REMOTE CONTROL
119
:DISPlay:DEMod[:WINDow]:TRACe:X
[:SCALe]:AUTO
Set
Query
Description
Turns auto scale on/off for AM/FM demodulation
or queries its state. When set to ON, the auto scale
function will be executed continuously.
Syntax
:DISPlay:DEMod[:WINDow]:TRACe:X[:SCALe]:AUTO
{OFF|ON|0|1}
Query Syntax
:DISPlay:DEMod[:WINDow]:TRACe:X[:SCALe]:AUTO?
Parameter
0
Turn Auto Scale off.
1
Turn Auto Scale on.
OFF
Turn Auto Scale off.
ON
Turn Auto Scale on.
Return parameter
0
Auto Scale is off.
1
Auto Scale is on.
Example
:DISP:DEM:TRAC:X:AUTO ON
:DISPlay:DEMod[:WINDow]:TRACe:X
[:SCALe]:PDIVision
Set
Query
Description
Sets or queries the time axis scale/div.
Syntax
:DISPlay:DEMod[:WINDow]:TRACe:X[:SCALe]:
PDIVision <time>
Query Syntax
:DISPlay:DEMod[:WINDow]:TRACe:X[:SCALe]:
PDIVision?
Parameter
<time>
<NRf>
Return parameter
<NR3>
Seconds
Example
:DISP:DEM:TRAC:X:PDIV 2 ms