User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
82
Adding a Step to the Automatic Test
Background
Up to 16 MANU tests (steps) can be added to
an automatic (AUTO) test. Each step is added
in a sequential order.
Steps
1. Press the DOWN arrow keys to
bring the cursor to the MANU
number.
I R G B
cursor
MANU number
m A
A CW D CW A D D
E
F R E Q = 0 H z 6
0
100
k V
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
1A U T O = 0 0 1 - 0 0
2. Use the scroll wheel to choose a
MANU number to add to the
automatic test.
MANU number
001~100
3. Press the ADD soft-key to add the
selected manual test to the
automatic test as another step.
A D D
4. Repeat steps 2 and 3 for any other tests that you
wish to add to the automatic test.
Note
After 16 steps have been added to an AUTO test,
FULL will be shown on the display when you
attempt to add another step to the AUTO test.