User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
OPERATION
79
Automatic Tests
This section describes how to create, edit and run automatic tests.
Automatic tests allow you to link together up to 16 different
MANU tests and run them sequentially. Each stored MANU test is
used as a test step when creating an AUTO test.
Choose/Recall an Automatic Test→ from page 79
Edit Automatic Test Settings → from page 81
Adding a Step to the Automatic Test → from page 82
Creating an AUTO Test File Name → from page 83
Saving and Exiting EDIT Status → from page 84
Automatic Test Page View → from page 85
Running an Automatic Test → from page 88
Automatic Test Results → from page 92
Before operating the GPT-9000 please read the safety precautions as
outlined in the Set Up chapter on page 22.
Choose/Recall an Automatic Test
Background
The tester must first be put into AUTO mode to
create or run automatic tests.
Up to 100 automatic tests can saved/recalled.
Steps
1. If the tester is in MANU mode,
press and hold the MANU/AUTO
key for three seconds. This will put
the tester into Auto mode.
The tester can only switch between
AUTO and MANU mode when in
the VIEW status.
MANU/AUTO