User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
68
IR PASS Timing
Discharge time
Output V
time
Initial time
(100ms)
START
TEST
PASS
RAMP TEST TIME
GB PASS Timing
Output I
time
START
TEST
PASS
0.5s
Initial time
(100ms)
Total Test Time
FAIL Judgment
1. When the test is judged as FAIL,
FAIL will be displayed, the buzzer
will sound and the FAIL indicator
will be lit red.
As soon as a test is judged FAIL,
power is cut from the terminals.
FAIL
I R G B A CW D CW
F R E Q = 0 H z 6
0
100
k V
IAF
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
2M A UN 0 0_2M A N U = * * * - 0 0
L
m A
01 37
2. The FAIL judgment will be held on
the display until the STOP button
is pressed. Pressing the STOP
button twice will return the tester
to the READY status.
STOP