GPT-9000 Series User Manual
52
Note
The ARC setting range is directly related to the HI
SET current limit.
ACW: GPT-98XX
HI SET Limit
ARC Range
0.001mA~0.999mA
1.000mA ~2.000mA
01.00mA~09.99mA
01.00mA ~20.00mA
010.0mA~042.0mA
001.0mA ~080.0mA
ACW: GPT-99XX
HI SET Limit
ARC Range
0.001mA~1.100mA
2.000mA
01.11mA~11.00mA
02.00mA ~20.00mA
011.1mA~110.0mA
002.0mA ~200.0mA
DCW: GPT-98XX
HI SET Limit
ARC Range
0.001mA~0.999mA
1.000mA ~2.000mA
01.00mA~09.99mA
01.00mA ~20.00mA
010.0mA~011.0mA
001.0mA ~020.0mA
DCW: GPT-99XX
HI SET Limit
ARC Range
0.001mA~1.100mA
2.000mA
01.11mA~11.00mA
02.00mA ~20.00mA
011.1mA~021.0mA
002.0mA ~040.0mA