GPT-9000 Series User Manual
36
GB Connection
Background
GB tests use the SENSE H/L and SOURCE H/L
terminals with the GTL-115 test leads.
GB Connection
Source H
DUT
GPT-9000
Sense L
Sense H
Source L
Steps
1. Turn the power off on the safety tester.
2. Connect the Sense H lead to the SENSE H
terminal.
3. Connect the Sense L lead to the SENSE L
terminal.
4. Connect the Source H lead to the SOURCE H
terminal.
5. Connect the Source L lead to the SOURCE L
terminal.
SOURCE H
SENSE H
SOURCE L
SENSE L
GB Rx
RETURN
HI-POT
IR
Sense H lead Sense L lead
Source L leadSource H lead