User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
20
Line voltage input
220
230
100
120
Line voltage input:
100/120/220/230VAC ±10%
Line voltage fuse
220
230
100
120
Line voltage selector and fuse:
GPT-98XX:
100V/120V
T5A 250V
220V/230V
T2.5A 250V
GPT-99XX:
100V/120V
T10A 250V
220V/230V
T6.3A 250V
Optional GPIB
port
GPIB
Optional GPIB interface for remote
control.
HIGH VOLTAGE
output terminal
GPT-9903/9904
CAUTION
HIGH VOLTAGE
MAX.
5.0 kVAC
6.0 kVDC
The HIGH VOLTAGE
terminal output is used for
outputting the testing
voltage.
WARNING
USE EXTREME CAUTION.
Do not touch the HIGH VOLTAGE terminal
during testing.
RETURN terminal
GPT-9903
RETURN
The RETURN terminal is
used for IR, DCW and ACW
tests.