- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
166
Table 1a: Output Limitation in Withstanding Voltage Testing (GPT-98XX)
Upper Current
Pause
Output Time
AC
30mA≤I≤40mA
At least as long
as the output
time
Maximum 240
seconds
0.001mA≤I<30
mA
Not necessary
Continuous
output possible
DC
0.001mA≤I≤10
mA
Not necessary
Continuous
output possible
GB
15A<I≤30A
(GPT-98XX)
15A<I≤32A
(GPT-99XX)
At least as long
as the output
time
999.9
3A≤I≤15A
Not necessary
999.9
NOTE: Output Time = Ramp Time + Test Time.
Table 1b: Output Limitation in Withstanding Voltage Testing (GPT-99XX)
Upper Current
Pause
Output Time
AC
80mA≤I≤100mA
At least as long
as the output
time
Maximum 240
seconds
0.001mA≤I<80
mA
Not necessary
Continuous
output possible
DC
0.001mA≤I≤20
mA
Not necessary
Continuous
output possible
GB
15A<I≤30A
(GPT-98XX)
15A<I≤32A
(GPT-99XX)
At least as long
as the output
time
999.9
3A≤I≤15A
Not necessary
999.9
NOTE: Output Time = Ramp Time + Test Time.