User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
APPENDIX
161
GPT-9000 Specifications
The specifications apply when the GPT-9000 is powered on for at
least 30 minutes at 15˚C~35˚C.
Specifications
Environment
Range
Temperature
Humidity
Warranty
15˚C ~ 35˚C
≤70% (No
condensation)
Operation
0˚C ~ 40˚C
≤70% (No
condensation)
Storage
-10˚C ~ 70˚C
≤85% (No
condensation)
Installation Location
Indoors at an amplitude of up to 2000m.
AC Withstanding Voltage
Output Voltage Range
0.100kV~ 5.000kV
Output Voltage Resolution
2V
Output Voltage Accuracy
(1% of setting +5V) with no load
Maximum Rated Load(Table1)
200 VA (5kV/40mA)[GPT-98XX]
500 VA (5kV/100mA)[GPT-99XX]
Maximum Rated Current
40mA [GPT-98XX], 100mA[GPT-99XX]
0.001mA ~ 10mA(0.1kV≤V≤0.5kV)
0.001mA ~ 40mA(0.5kV<V≤5kV) [GPT-98XX]
0.001mA ~ 100mA(0.5kV<V≤5kV) [GPT-99XX]
Output Voltage Waveform
Sine wave
Frequency
50 Hz / 60 Hz
Voltage Regulation
1% +5V
[Maximum rated load → no load]
Voltmeter Accuracy
(1% of reading+ 5V)
Current Measurement Range
0.001mA~040.0mA [GPT-98XX]
0.001mA~100.0mA [GPT-99XX]
Continued…