User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
151
AUTO:PAGE:DEL
Set
Query
Description
Deletes the selected step from the AUTO test. The
remaining steps move up to replace the deleted
step.
Syntax
AUTO:PAGE:DEL <NR1>
Parameter/
<NR1>
1~16 (step no.#)
Example
AUTO:PAGE:DEL 3
Deletes the contents of step number #3.
A CW = 0 . 0 0 k1
O V EM W A PS K I PS D E L
E MA U OT N A_ 0A U T O = 0 0 1 - 0 1
M A N U _ N M E A V H
# 0 1 : 0 1 0 # 0 2 : 0 0 1 # 0 3 : 0 0 3 # 0 4 : 0 0 4
# 0 5 : 0 0 7 # 0 6 : 0 0 3 # 0 7 : 0 3 8 # 0 8 : 0 0 5
# 0 9 : # 1 0 : # 1 1 : # 1 2 :
# 1 3 : # 1 4 : # 1 5 : # 1 6 :
1 . 0 0 m AES = 0T_I
AUTO:NAME
Set
Query
Description
Sets or returns the AUTO name for the selected
automatic test. The test must be in AUTO mode
before this command can be used.
Note only alphanumeric characters (A-Z, a-z, 0-9)
and the “_” underscore character can be used to set
the AUTO test name.
Syntax
Query Syntax
AUTO:NAME <string>
AUTO:NAME?
Parameter/
Return parameter
<string>
10 character string. (first character must
be a letter)
Example
AUTO:NAME program1
Sets the AUTO name to “program1”.