User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
146
All Data Points
Example
SWEEP0:DATA:SHOW?
>ACW MODE,CR+LF
>No.,V(kV),I(mA), T(S) ,CR+LF
>001,0.071,0.032,0000.1,CR+LF
>002,0.111,0.047,0000.2,CR+LF
>…………………………………
>013,0.601,0.215,0001.3,CR+LF
>END
This will return all the data from the sweep graph.
SWEEP:GRAPh:SHOW
Set
Query
Description
Turns the sweep graph on or off on the GPT-99XX
display.
Syntax
Query Syntax
SWEEP:GRAPh:SHOW {ON|OFF}
SWEEP:GRAPh:SHOW?
Parameter/
Return parameter
ON
Turn the sweep graph on.
OFF
Turn the sweep graph off.
Example
SWEEP:GRAP:SHOW ON
Displays the sweep graph on the LCD display.
SWEEP :GRAPh:LINE
Set
Query
Description
Sets or returns which lines are shown on the sweep
graph.
Syntax
Query Syntax
SWEEP:GRAPh:LINE <NR1>
SWEEP:GRAPh:LINE?