User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
141
MANU:UTILity:ARCMode
Set
Query
Description
Sets or returns the ARC mode status for the
current test.
The ARC mode cannot be set for the IR and GB
function.
Syntax
Query Syntax
MANU:UTILity:ARCMode {OFF|ON_CONT|
ON_STOP}
MANU:UTILity:ARCMode?
Parameter/
Return parameter
OFF
Turns ARC mode off.
ON_CONT
Sets ARC mode to ON and
CONTINUE.
ON_STOP
Sets ARC mode to ON and STOP.
Example
MANU:UTIL:ARCM OFF
Turns ARC mode OFF.
MANU:UTILity:PASShold
Set
Query
Description
Sets or returns the PASS HOLD setting for the
current test.
Syntax
Query Syntax
MANU:UTILity:PASShold {ON|OFF}
MANU:UTILity:PASShold?
Parameter/
Return parameter
OFF
Turns PASS HOLD off.
ON
Turns PASS HOLD on.
Example
MANU:UTIL:PASS OFF
Turns PASS HOLD OFF.
MANU:UTILity:FAILmode
Set
Query
Description
Sets or returns the FAIL mode setting for the
current test.
Syntax
Query Syntax
MANU:UTILity:FAILmode {CONT|HOLD|STOP}
MANU:UTILity:FAILmode?