User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
134
Syntax
Query Syntax
MANU:DCW:CLOSet<NR2>
MANU:DCW:CLOSet?
Parameter/
Return parameter
<NR2>
0.000 ~ 010.9 (GPT-98XX)
0.000 ~ 020.9 (GPT-99XX)
Example
MANU:DCW:CLOS 2.00
Sets the DCW LO SET current to 2mA.
MANU:DCW:TTIMe
Set
Query
Description
Sets or returns the DCW test time in seconds. The
test must first be in DCW mode before this
command can be used.
In special MANU mode, the TIMER can be turned
off.
Syntax
Query Syntax
MANU:DCW:TTIMe {<NR2>|OFF}
MANU:DCW:TTIMe?
Parameter
<NR2>
OFF
0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).
Return parameter
<NR2>
TIME OFF
0.5 ~ 999.9 seconds
TIMER is OFF (special MANU mode).
Example
MANU:DCW:TTIM 1
Sets the DCW test time to 1 second.
MANU:DCW:REF
Set
Query
Description
Sets or returns the DCW reference value in mA.
The test must first be in DCW mode before this
command can be used.
The reference value must be less than the HI SET
value.
The reference value uses the same range as the HI
SET value.