User manual

Table Of Contents
GPT-9000 Series User Manual
134
MANU:DCW:CLOSet<NR2>
MANU:DCW:CLOSet?
<NR2>
0.000 ~ 010.9 (GPT-98XX)
0.000 ~ 020.9 (GPT-99XX)
MANU:DCW:CLOS 2.00
Sets the DCW LO SET current to 2mA.
MANU:DCW:TTIMe
Set
Query
Sets or returns the DCW test time in seconds. The
test must first be in DCW mode before this
command can be used.
In special MANU mode, the TIMER can be turned
off.
MANU:DCW:TTIMe {<NR2>|OFF}
MANU:DCW:TTIMe?
<NR2>
OFF
0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).
<NR2>
TIME OFF
0.5 ~ 999.9 seconds
TIMER is OFF (special MANU mode).
MANU:DCW:TTIM 1
Sets the DCW test time to 1 second.
MANU:DCW:REF
Set
Query
Sets or returns the DCW reference value in mA.
The test must first be in DCW mode before this
command can be used.
The reference value must be less than the HI SET
value.
The reference value uses the same range as the HI
SET value.