REMOTE CONTROL
133
MANU:DCW:CHISet
Set
Query
Sets or returns the DCW HI SET current value in
milliamps. The test must first be in DCW mode
before this command can be used.
Note: A “DC Over 50W” error will result if the
DCW Voltage X HI SET value is > 50 watts.
MANU:DCW:CHISet <NR2>
MANU:DCW:CHISet?
<NR2>
0.001 ~ 011.0 (GPT-98XX)
0.001 ~ 021.0 (GPT-99XX)
MANU:DCW:CHIS 5
Sets the DCW HI SET current to 5mA.
MANU:DCW:CLOSet
Set
Query
Sets or returns the DCW LO SET current value in
milliamps. The LO SET value must be less than the
HI SET value. The test must first be in DCW mode
before this command can be used.
The LO SET range must use the HI SET range. If all
the digits in the LO SET range are outside the HI
SET range, an error will be produced. All digits
outside the HI SET range are ignored and will not
be used.
For example:
HI SET value: 12.34
LO SET value1: 0.005 error
LO SET value2: 0.053 no error
In the example above LO SET value1 will produce
an error as all digits are outside the range of HI
SET. LO SET value2 will not produce an error, but
will return 0.05, not 0.053.