User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
132
MANU:ACW:ARCCurrent
Set
Query
Description
Sets or returns the ACW ARC current value in mA.
ARC must be enabled before the ARC current can
be set. The test must first be in ACW mode before
this command can be used.
ARC current uses the same range as the HI SET
value. The ARC current is limited to 2X the HI SET
value.
Syntax
Query Syntax
MANU:ACW:ARCCurrent <NR2>
MANU:ACW:ARCCurrent?
Parameter/
Return parameter
<NR2>
1.000 ~ 080.0 (GPT-98XX)
2.000 ~ 200.0 (GPT-99XX)
Example
MANU:ACW:ARCC 0.04
Sets the ACW ARC value to 0.04 mA.
MANU:DCW:VOLTage
Set
Query
Description
Sets or returns the DCW voltage in kV. The test
must first be in DCW mode before this command
can be used.
Note: A “DC Over 50W” error will result if the
DCW Voltage X HI SET value is > 50 watts.
Syntax
Query Syntax
MANU:DCW:VOLTage <NR2>
MANU:DCW:VOLTage?
Parameter/
Return parameter
<NR2>
0.100 ~ 6.100 (kV)
Example
MANU:DCW:VOLT 6
Sets the DCW voltage to 6 kV.