User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
129
MANU:ACW:CHISet
Set
Query
Description
Sets or returns the ACW HI SET current value in
milliamps. The test must first be in ACW mode
before this command can be used.
Syntax
Query Syntax
MANU:ACW:CHISet <NR2>
MANU:ACW:CHISet?
Parameter/
Return parameter
<NR2>
0.001 ~ 042.0 (GPT-98XX)
0.001 ~ 110.0 (GPT-99XX)
Example
MANU:ACW:CHIS 10.0
Sets the ACW HI SET current to 10 mA.
MANU:ACW:CLOSet
Set
Query
Description
Sets or returns the ACW LO SET current value in
milliamps. The LO SET value must be less than the
HI SET value. The test must first be in ACW mode
before this command can be used.
The LO SET range must use the HI SET range. If all
the digits in the LO SET range are outside the HI
SET range, an error will be produced. All digits
outside the HI SET range are ignored and will not
be used.
For example:
HI SET value: 12.34
LO SET value1: 0.005 error
LO SET value2: 0.053 no error
In the example above LO SET value1 will produce
an error as all digits are outside the range of HI
SET. LO SET value2 will not produce an error, but
will return 0.05, not 0.053.