User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
OPERATION
75
Test V
Test I
time
Start time
RAMP TIME
Legend: Voltage: Current
TEST time
Voltage
Current
The test items that are plotted on the sweep
graph depend on the type of test that is
performed.
TEST
Graph Test Items
ACW
Test voltage, test current (V, I)
DCW
Test voltage, test current (V, I)
IR
Test current, test resistance (I, R)
GB
Test voltage, test resistance (V, R)
Steps
1. Choose MANU number 000 to
enter the special test mode.
Page 38
2. The settings of a previous test can
be loaded by pressing the
corresponding soft-key in the
VIEW or READY status.
For example, if you are currently in
DCW mode, pressing the ACW
key will load the ACW settings
that were previously used in the
special manual mode.
A CW
Example:
ACW