User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
72
4. The ZERO function can be
activated by pressing the
corresponding soft-key in the
READY status. The ZERO soft-key
will be highlighted.
E R OZ
5. Press the START button to perform
the zero check. The tester will go
into the ZERO status.
START
0
3 01
A
Z E R 0 I R G B A CW D CW
E
F R E Q = 0 H z 6
REZ
G B V = 0 . 6 2 2 V MT I E = 0 0 3 . 2 SR
. 0 m ΩER F =# 0 00
0 0 . 0 m ΩIH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
O
Measured Resistance
200 7
m Ω
ZERO status
6. When the zero check has finished, the tester
will return back to the VIEW status. The
resistance of the test leads will be automatically
set as the Reference value.
0
3 01
A
I R G B A CW D CW
E
F R E Q = 0 H z 6
EIV
G B V = V MT I E = 0 0 3 . 2 SR
. 7 m ΩER F =# 0 02
0 0 . 0 m ΩIH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
W
m Ω
Reference value
Note
Remember to replace the test leads to the proper
position on the DUT before testing.