User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
OPERATION
71
Zeroing of the Test Leads (GB only)
Background
The Zeroing function is used to determine the
resistance of the test leads for GB tests. When a
zero check is performed, the reference is
automatically set to the measured resistance of
the test leads.
This function is only available for GB testing.
Steps
1. Ensure the tester is in VIEW status
for the current GB test. Save the
current test if necessary.
Page 61
I R G B A CW D CW
E
F R E Q = 0 H z 6
EIV
G B V = V MT I E = 0 0 3 . 2 SR
. 0 m ΩER F =# 0 00
0 0 . 0 m ΩIH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
W
m Ω
0
3 01
A
2. Short the positive and negative alligator clips as
shown below.
3. Press the STOP button to put the
tester into the READY status.
STOP