User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
OPERATION
55
2. Use the UP/DOWN arrow keys to
move to the FAIL MODE setting.
3. Use the scroll wheel to set FAIL
MODE.
FAIL MODE
CONTINUE, HOLD, STOP
4. Press the EDIT/SAVE key to save
and exit the MANU Utility menu.
EDIT/SAVE
Note
The ESC key can be pressed at any time in the
MANU Utility menu to cancel and exit.
Setting MAX HOLD
Background
The MAX HOLD setting will hold the
maximum current measured in the ACW and
DCW tests or the maximum resistance
measured in IR and GB tests.
Steps
1. Press the UTILITY key on the front
panel when the tester is in EDIT
status. The display will go from the
normal EDIT status to the MANU
Utility menu for the current test.
UTILITY
L I T Y IM A UN U T 2M A N U = * * * - 0 0
A R C M D E : OO FF
G R O U N D MO
P A S S H L D :O O FF
F A I L M D E :O
PS T O
M A X H O L D : O F F
D E : NO
Note
The MANU UTILITY settings only apply to the
selected MANU test.