User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
54
Note
The ESC key can be pressed at any time in the
MANU Utility menu to cancel and exit.
Setting FAIL MODE
Background
The FAIL MODE settings only apply to the
selected test in AUTO tests.
FAIL MODE has three options, CONTINUE,
HOLD and STOP.
When FAIL MODE is set to CONTINUE the
tester will continue testing after a FAIL
judgment.
When set to HOLD, the tester will hold the test
on a FAIL judgment, and then continue testing
after the START key is pressed.
The STOP mode will completely stop the test
after a FAIL judgment.
Note
The FAIL MODE setting only applies to AUTO
tests. This setting is ignored when running
MANU tests.
Steps
1. Press the UTILITY key on the front
panel when the tester is in
MANU/EDIT status. The display
will go from the normal EDIT
status to the MANU Utility menu
for the current test.
UTILITY
L I T Y IM A UN U T 2M A N U = * * * - 0 0
A R C M D E : OO FF
G R O U N D M O D E : NO
P A S S H L D :O O FF
F A I L M D E :O
PS T O
M A X H O L D : O F F