User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
40
Setting the Test Function
Background
After a MANU number has been chosen and
the tester is in EDIT status, a test function can
be set.
There are four test functions, AC Withstand,
DC Withstand, Insulation Resistance and
Ground Bond.
Steps
1. To choose the test function, press the ACW,
DCW, IR or GB soft-keys.
A CW D CW I R
G B
2. The test function soft-key is highlighted.
I R G B A CW D CW I / LH O I M ET R
E
F R E Q = 0 H z 6
0
100
k V
IDE
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
test function
m A
Note
The chosen test function only applies to the
current test.
Setting the Test Voltage or Test Current
Background
The test voltage can be set from 0.100kV to 5kV
for ACW, 0.100kV to 6kV for DCW and 0.050 to
1kV for IR (50V steps). For GB tests the test
current can be set from 3A to 30A (GPT-98XX)
or 3A to 32A (GPT-99XX).