User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
152
AUTO:EDIT:ADD
Set
Query
Description
Add the selected MANU test to the current AUTO
number.
Syntax
AUTO:EDIT:ADD <NR1>
Parameter/
<NR1>
1~100
Example
AUTO:EDIT:ADD 7
Adds MANU-007 to the current AUTO number. I.e.,
A CW = 0 . 0 0 k1
O V EM W A PS K I PS D E L
E MA U OT N A_ 7A U T O = 0 0 5 - 0 0
M A N U _ N M E A V H
# 0 1 : 0 1 0 # 0 2 : 0 0 1 # 0 3 : 0 0 3 # 0 4 : 0 0 4
# 0 5 : 0 0 7 # 0 6 : # 0 7 : # 0 8 :
# 0 9 : # 1 0 : # 1 1 : # 1 2 :
# 1 3 : # 1 4 : # 1 5 : # 1 6 :
MANU test added to
last step
1 . 0 0 m AES = 0T_I
TESTok:RETurn
Set
Query
Description
Allows “OK” to be displayed on the remote
terminal when a test has stopped (PASS/FAIL or
STOP). This applies for MANU and AUTO mode.
By default, TESTok:RETurn is set to OFF.
Syntax
Query Syntax
TESTok:RETurn {ON|OFF}
TESTok:RETurn?
Parameter/
Return parameter
ON
Enables the “OK” message to be
displayed.
OFF
Disables the message
Example
TEST:RET OFF
Disables the message.