User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
139
Syntax
Query Syntax
MANU:GB:RLOSet<NR2>
MANU:IR:RLOSet?
Parameter/
Return parameter
<NR2>
0.000 ~ 649.9
Example
MANU:GB:RLOS 50
Sets the GB LO SET resistance to 50mΩ.
MANU:GB:TTIMe
Set
Query
Description
Sets or returns the GB test time in seconds. The test
must first be in GB mode before this command can
be used.
Syntax
Query Syntax
MANU:GB:TTIMe <NR2>
MANU:GB:TTIMe?
Parameter/
Return parameter
<NR2>
0.5 ~ 999.9 seconds
Example
MANU:GB:TTIM 1
Sets the GB test time to 1 second.
MANU:GB:FREQuency
Set
Query
Description
Sets or returns the GB test frequency in Hz. The
test must first be in GB mode before this command
can be used.
Syntax
Query Syntax
MANU:GB:FREQuency {50|60}
MANU:GB:FREQuency?
Parameter/
Return parameter
<50>
50 Hz
<60>
60 Hz
Example
MANU:GB:FREQ 50
Sets the GB test frequency to 50Hz.