User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
138
MANU:GB:CURRent
Set
Query
Description
Sets or returns the GB current in A. The test must
first be in GB mode before this command can be
used.
Syntax
Query Syntax
MANU:GB:CURRent <NR2>
MANU:GB:CURRent?
Parameter/
Return parameter
<NR2>
3.00~32.00 (GPT-98XX)
3.00~33.00 (GPT-99XX)
Example
MANU:GB:CURR 3.00
Sets the GB current to 3.00A.
MANU:GB:RHISet
Set
Query
Description
Sets or returns the GB HI SET resistance value in
mΩ. The test must first be in GB mode before this
command can be used.
Syntax
Query Syntax
MANU:GB:RHISet <NR2>
MANU:GB:RHISet?
Parameter/
Return parameter
<NR2>
000.1 ~ 650.0
Example
MANU:GB:RHIS 100.0
Sets the HI SET value to 100mΩ.
Note
If the (GB current x HI SET resistance) > 5.4V, then an
error will be generated (“GBV > 5.4V”).
MANU:GB:RLOSet
Set
Query
Description
Sets or returns the GB LO SET resistance value in
mΩ. The LO SET value must be less than the HI
SET value. The test must first be in GB mode
before this command can be used.