User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
136
Example
MANU:IR:VOLT 1
Sets the IR voltage to 1 kV.
MANU:IR:RHISet
Set
Query
Description
Sets or returns the IR HI SET resistance value in
MΩ (GPT-98XX) or GΩ (GPT-99XX). The test must
first be in IR mode before this command can be
used.
Syntax
Query Syntax
MANU:IR:RHISet <NR1>|NULL
MANU:IR:RHISet?
Parameter/
Return parameter
<NR1>
2 ~ 9999 (GPT-98XX)
0.002 ~ 50.00 (GPT-99XX)
NULL
Sets the HI SET value to high impedance
Example
(GPT-98XX)
MANU:IR:RHIS 10
Sets the IR HI SET resistance to 10 MΩ.
Example
(GPT-99XX)
MANU:IR:RHIS 0.010
Sets the IR HI SET resistance to 10 MΩ.
MANU:IR:RLOSet
Set
Query
Description
Sets or returns the IR LO SET resistance value in
MΩ (GPT-98XX) or GΩ (GPT-99XX). The LO SET
value must be less than the HI SET value. The test
must first be in IR mode before this command can
be used.
Syntax
Query Syntax
MANU:IR:RLOSet<NR1>
MANU:IR:RLOSet?
Parameter/
Return parameter
<NR1>
1 ~ 9999 (GPT-98XX)
0.001 ~ 50.00 (GPT-99XX)
Example
(GPT-98XX)
MANU:IR:RLOS 10
Sets the IR LO SET resistance to 10MΩ.