GPT-9000 Series User Manual
132
MANU:ACW:ARCCurrent
Set
Query
Sets or returns the ACW ARC current value in mA.
ARC must be enabled before the ARC current can
be set. The test must first be in ACW mode before
this command can be used.
ARC current uses the same range as the HI SET
value. The ARC current is limited to 2X the HI SET
value.
MANU:ACW:ARCCurrent <NR2>
MANU:ACW:ARCCurrent?
<NR2>
1.000 ~ 080.0 (GPT-98XX)
2.000 ~ 200.0 (GPT-99XX)
MANU:ACW:ARCC 0.04
Sets the ACW ARC value to 0.04 mA.
MANU:DCW:VOLTage
Set
Query
Sets or returns the DCW voltage in kV. The test
must first be in DCW mode before this command
can be used.
Note: A “DC Over 50W” error will result if the
DCW Voltage X HI SET value is > 50 watts.
MANU:DCW:VOLTage <NR2>
MANU:DCW:VOLTage?
<NR2>
0.100 ~ 6.100 (kV)
MANU:DCW:VOLT 6
Sets the DCW voltage to 6 kV.