User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
GPT-9000 Series User Manual
128
Syntax
Query Syntax
MANU:RTIMe <NR2>
MANU:RTIMe?
Parameter/
Return parameter
<NR2>
0.1~999.9 seconds
Example
MANU:RTIM 0.5
Sets the ramp time to half a second.
MANU:EDIT:MODE
Set
Query
Description
Sets or returns the mode (ACW, DCW, IR) of the
selected manual test.
Syntax
Query Syntax
MANU:EDIT:MODE {ACW|DCW|IR|GB}
MANU:EDIT:MODE?
Parameter/
Return parameter
<ACW>
AC Withstand mode
<DCW>
DC Withstand mode
<IR>
Insulation Resistance mode
Example
MANU:EDIT:MODE ACW
Sets the mode to ACW.
MANU:ACW:VOLTage
Set
Query
Description
Sets or returns the ACW voltage in kV. The test
must first be in ACW mode before this command
can be used.
Syntax
Query Syntax
MANU:ACW:VOLTage <NR2>
MANU:ACW:VOLTage?
Parameter/
Return parameter
<NR2>
0.100 ~ 5.000 (kV)
Example
MANU:ACW:VOLT 1
Sets the ACW voltage to 1 kV.