User manual

Table Of Contents
REMOTE CONTROL
127
MANU:STEP
Set
Query
Sets the MANU test number.
MANU:STEP <NR1>
MANU:STEP?
<NR1>
0~100.
MANU:STEP 100
Sets the manual test number to 100.
MANU:NAME
Set
Query
Sets or returns the test name for the selected
manual test. The test must be in MANU mode
before this command can be used.
Note only alphanumeric characters (A-Z, a-z, 0-9)
and the “_” underscore character can be used to set
the MANU test name.
MANU:NAME <string>
MANU:NAME?
<string>
10 character string. (first character must
be a letter)
MANU:NAME test1
Sets the manual test name to “test1”.
MANU:RTIMe
Set
Query
Sets or returns the Ramp Time for the test in
seconds.
Note: A “TIME ERR” will result if the Ramp Time
+ Test Time is ≥ 240 seconds when the HI SET limit
is over 30mA (GPT-98XX) or over 80mA (GPT-
99XX). This applies to the ACW function only.