User manual
Table Of Contents
- SAFETY INSTRUCTIONS
- GETTING STARTED
- OPERATION
- Menu Tree
- Test Lead Connection
- ACW, DCW and GB Manual Testing
- Choose/Recall a Manual Test Number
- Edit Manual Test Settings
- Setting the Test Function
- Setting the Test Voltage or Test Current
- Setting the Test Frequency
- Setting the Upper and Lower Limits
- Setting a Reference Value
- Setting the Test Time (Timer)
- Setting the Ramp Up Time
- Creating a MANU Test File Name
- Setting the ARC Mode
- Setting PASS HOLD
- Setting FAIL MODE
- Setting MAX HOLD
- Setting the Grounding Mode
- Saving and Exiting EDIT Status
- Running a MANU Test
- PASS / FAIL MANU Test
- Zeroing of the Test Leads (GB only)
- Special MANU Test Mode (000)
- Automatic Tests
- Common Utility Settings
- EXTERNAL CONTROL
- REMOTE CONTROL
- FAQ
- APPENDIX
- INDEX
REMOTE CONTROL
127
MANU:STEP
Set
Query
Description
Sets the MANU test number.
Syntax
Query Syntax
MANU:STEP <NR1>
MANU:STEP?
Parameter/
Return parameter
<NR1>
0~100.
Example
MANU:STEP 100
Sets the manual test number to 100.
MANU:NAME
Set
Query
Description
Sets or returns the test name for the selected
manual test. The test must be in MANU mode
before this command can be used.
Note only alphanumeric characters (A-Z, a-z, 0-9)
and the “_” underscore character can be used to set
the MANU test name.
Syntax
Query Syntax
MANU:NAME <string>
MANU:NAME?
Parameter/
Return parameter
<string>
10 character string. (first character must
be a letter)
Example
MANU:NAME test1
Sets the manual test name to “test1”.
MANU:RTIMe
Set
Query
Description
Sets or returns the Ramp Time for the test in
seconds.
Note: A “TIME ERR” will result if the Ramp Time
+ Test Time is ≥ 240 seconds when the HI SET limit
is over 30mA (GPT-98XX) or over 80mA (GPT-
99XX). This applies to the ACW function only.